Calibration Security; Unsecuring The Instrument For Calibration - Agilent Technologies U1253A User's And Service Manual

True rms oled multimeter
Hide thumbs Also See for U1253A:
Table of Contents

Advertisement

Calibration Security

N O T E
N O T E

Unsecuring the instrument for calibration

U1253A User's and Service Guide
A calibration security code is in place to prevent accidental
or unauthorized adjustments to the U1253A true RMS OLED
multimeter. When you first receive your instrument, it is
secured. Before you can adjust the instrument, you must
"unsecure" it by entering the correct security code (see
"Unsecuring the instrument for calibration"
The security code is set as 1234 when the instrument is
shipped from the factory. The security code is stored in
nonvolatile memory, which does not change even when the
power is off.
You can unsecure the instrument and then change the security code from
the front panel or through the remote interface.
See
"Resetting the security code to factory default"
forget your security code.
Before you can adjust the instrument, you must unsecure it
by entering the correct security code, either from the front
panel, or through PC remote interface.
The default security code is 1234.
From front panel
1 Turn the rotary switch to the
also start with another rotary switch position; but here
we assume that you will follow the exact steps listed in
Table 6-
2).
2 Press
and
Calibration Security Code entry mode.
Performance Tests and Calibration
on page 147).
on page 152 if you
position (you may
simultaneously to enter the
6
147

Advertisement

Table of Contents
loading

Table of Contents