Agilent MOI for MHL Cables Compliance Tests
6. [Appendix] Manual Setup
The procedures of manual setup for time-domain and frequency-domain measurements are
introduced in the section. All the following parameters are saved in the E5071C's state file,
which is available at:
6.1. Channel & Trace Setup
If TDR setup wizard is shown when launching the TDR software, click
TDR setup wizard main window.
1. Open
Setup
tab in the TDR software.
2. Click
Preset
to preset the instrument. Click
3. Set
DUT Topology
4. Click
Advanced
5. A dialog box appears requesting for confirmation. Then click Yes. (Clear the check box
for "Use Advanced Calibration Methods")
6. Click
Stop
Single.
7. Open
TDR/TDT
8. Click
Trace Control
9. Clear
Time
and
www.agilent.com/find/ena-tdr_mhl-cabcon
to "Single-Ended 4-Port". Click
Mode>>.
tab.
tab.
Marker
check box under Coupling.
OK
in a dialog box to continue.
OK
in a dialog box.
26
Close
button in the