Table 6-2 Performance Verification Tests (Continued) - Agilent Technologies U1211A User's Manual And Service Manual

Clamp meters
Hide thumbs Also See for U1211A:
Table of Contents

Advertisement

Table 6-2 Performance verification tests (continued)

Test function
[1]
DC current
AC current
[2]
AC + DC current
Peak current (max)
[1] The measurement option is only available for U1212A and U1213A.
[2] The measurement option is only available for U1213A.
User's and Service Guide
Range
5520A output
use with
5500A/COIL
40 A
0.8 A
400 A
8 A
1000 A
20 A
40 A
0.8 A, 45 Hz
0.8 A, 100 Hz
0.8 A, 400 Hz
400 A
8 A, 45 Hz
0.4 A, 400 Hz
1000 A
14 A, 45 Hz
2.99999 A,
400 Hz
20 A, 45 Hz
40 A
0.8 A, 400 Hz
400 A
0.4 A, 400 Hz
1000 A
1 A, 400 Hz
400 A
8 A
,
peak
60 Hz
Performance Tests and Calibration
Performance Verification Tests
Reference
Error limits (from nominal 1 year)
values
U1211A
40 A
400 A
1000 A
40 A, 45 Hz
±0.5 A
40 A, 100 Hz
±0.5 A
40 A, 400 Hz
±0.5 A
400 A, 45 Hz
±4.5 A
20 A, 400 Hz
±0.7 A
700 A, 45 Hz
±12 A
150 A, 400 Hz
±6 A
1000 A, 45 Hz
40 A, 400 Hz
20 A, 400 Hz
50 A, 400 Hz
400 A
,
±12.3 A
peak
60 Hz
6
U1212A
U1213A
±0.75 A
±0.75 A
±6.5 A
±6.5 A
±25 A
±25 A
±0.9 A
±0.9 A
±1.3 A
±1.3 A
±1.3 A
±8.5 A
±8.5 A
±1.1 A
±1.1 A
±9 A
±9 A
±30 A
±30 A
±2.05 A
±1.8 A
±12 A
±12.3 A
±12.3 A
93