AC specifications

The AC voltage and AC current specifications are true RMS AC coupled,
valid from 5% to 100% of range. The crest factor may be up to 3 at full
scale, except for the 1000 V and 1000 A ranges which is 1.5 at full scale.
For non- sinusoidal waveforms with crest factor ≤ 3, add 2% reading + 2%
full scale typical.
Table 7-10 U1212A AC accuracy ± (% of reading + number of LSD)
Function
[1]
AC voltage
Function
[2]
AC current
[1]
[2]
User's and Service Guide
Range
400 V
1000 V
Range
Resolution
40 A
0.01 A
400 A
0.1 A
1000 A
1 A
Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
The maximum verification of current and frequency product is less than 400,000 A × Hz.
Characteristics and Specifications
Resolution
Accuracy
45 Hz to 400 Hz
0.1 V
1 V
Accuracy
45 Hz to 65 Hz
2.0% + 10
2.0% + 5
2.5% + 5
U1212A Electrical Specifications
Overload protection
1.0% + 5
1000 V
1.0% + 5
65 Hz to 1 kHz
3.0% + 10
3.0% + 5
3.0% + 5
7
rms
Overload
protection
1000 V
rms