Table 6-2 Performance Verification Tests - Agilent Technologies U1211A User's Manual And Service Manual

Clamp meters
Hide thumbs Also See for U1211A:
Table of Contents

Advertisement

Table 6-2 Performance verification tests

Test function
Temperature
Resistance
Diode
Capacitance
DC voltage
User's and Service Guide
Range
signals/values
5520A output
–200 °C to –40 °C
–40 °C to 1372 °C
–40 °C to 1372 °C
400 Ω
4 kΩ
40 kΩ
400 kΩ
4 MΩ
40 MΩ
Diode
4 µF
40 µF
400 µF
4000 µF
4 V
40 V
400 V
1000 V
Performance Tests and Calibration
Reference
Error limits (from nominal 1 year)
U1211A
–200 °C
0 °C
1372 °C
400 Ω
±2.3 Ω
4 kΩ
±0.023 kΩ
40 kΩ
400 kΩ
4 MΩ
40 MΩ
1.9 V
±0.012 V
4 µF
40 µF
400 µF
±8.4 µF
4000 µF
±124 µF
4 V
40 V
400 V
±2.3 V
1000 V
±8 V
Performance Verification Tests
U1212A
±5.0 °C
±1.0 °C
±14.7 °C
±2.3 Ω
±0.024 kΩ
±0.012 V
±8.4 µF
±124 µF
±2.3 V
±8 V
6
U1213A
±5.0 °C
±1.0 °C
±14.7 °C
±1.5 Ω
±0.015 kΩ
±0.15 kΩ
±1.5 kΩ
±0.027 MΩ
±0.85 MΩ
±0.012 V
±0.044 µF
±0.44 µF
±8.4 µF
±124 µF
±0.011 V
±0.11 V
±1.1 V
±8 V
91