Front Panel Calibration
Calibration process
The following general procedure is the recommended method to complete
a full instrument calibration:
1 Read
2 Perform the verification tests to characterize the instrument (incoming
3 Unsecure the instrument calibration (see
4 Perform the adjustment procedures (see
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the instrument
N O T E
Adjustment procedures
The calibration procedures are shown as follows:
1 Press Range/Auto for more than 1 second while rotating the rotary
2 Unsecure the clamp meter. Refer to
3 After verifying that the security code entered is correct, the instrument
4 Set up the indicated reference input and apply this input to the correct
User's and Service Guide
"Test Considerations"
data).
page 98).
page 101).
maintenance records.
Make sure that you quit the adjustment mode before turning off the
clamp meter.
switch to the function you wish to adjust.
calibration"
on page 98.
will indicate the reference input value of the adjustment item on the
primary display after briefly showing "PASS" on the secondary display.
terminals of the clamp meter. For example:
Performance Tests and Calibration
on page 88.
"Calibration Security"
"Adjustment Considerations"
"Unsecuring the instrument for
Front Panel Calibration
on
6
on
107