AC specifications

The AC voltage and AC current specifications are true RMS AC coupled,
valid from 5% to 100% of range. The crest factor may be up to 3 at full
scale, except for the 1000 V and 1000 A ranges which is 1.5 at full scale.
For non- sinusoidal waveforms with crest factor ≤ 3, add 2% reading + 2%
full scale typical.
Table 7-3 U1211A AC accuracy ± (% of reading + number of LSD)
Function
[1]
AC voltage
Function
[2]
AC current
400 A to 700 A
700 A to 1000 A
[1]
[2]
[3]
[4]
User's and Service Guide
Use the null operation to zero out residual offset before measuring the signal (by shorting the test
leads).
Range
400 V
1000 V
Range
Resolution
40 A
0.01 A
400 A
0.1 A
1 A
1 A
Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
Maximum overload: 1000 A
The accuracy for AC is specified on the symmetrical waveforms.
The maximum verification of current and frequency product is less than 400,000 A × Hz.
Characteristics and Specifications
Resolution
45 Hz to 400 Hz
0.1 V
1 V
45 Hz to 65 Hz
1.0% + 10
1.0% + 5
1.0% + 5
1.0% + 5
rms
U1211A Electrical Specifications
Accuracy
Overload protection
1.0% + 5
1000 V
1.0% + 5
[3][4]
Accuracy
65 Hz to 400 Hz
400 Hz to 1 kHz
1.0% + 10
1.0% + 5
1.0% + 5
7
rms
3.0% + 10
3.0% + 5
3.0% + 5