6
Performance Tests and Calibration
Adjustment Considerations
Table 6-6 U1212A valid adjustment reference input values (continued)
Function
Range
Temperature
Short
0.4 V
K-type
Diode
Short
2.000 V
Table 6-7 U1213A valid adjustment reference input values
Function
Range
DC voltage
Short
4 V
40 V
400 V
1000 V
104
Reference input value
SHORT
0.400 V
000.0 °C
SHORT
2.000 V
Reference input value
SHORT
3.000 V
30.00 V
300.0 V
1000 V
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
Provide 0 °C with ambient compensation
0 Ω
0.9 to 1.1 × reference input value
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
User's and Service Guide