Maintenance
7
Malfunctions and troubleshooting
7.2
Malfunctions and troubleshooting
The following tables describe potential faults and their elimination.
No.
Occurrence
1 The arm is difficult to posi-
tion
2 Arm does not stay in the po-
sition.
3 Hissing sound as a result of
pressure leakage
4 Distributor module cannot
be turned.
5 The distributor module col-
lides against other objects.
6 Splintering paint
7 LED does not light up.
8 LED on the side of the
drawer is weak or does not
light up at all.
86 / 104
Possible cause
Distributor module is over-
loaded
Brakes cannot be (fully) re-
leased.
The brakes are damaged.
The electro-pneumatic
brakes or electro-magnetic
brakes have an electrical
defect.
Mechanical brakes are worn
or not properly adjusted.
The pneumatic brakes are
damaged.
The base plate of the spacer
tube is not horizontally
aligned.
The pneumatic brakes are
damaged.
Distributor module is over-
loaded.
Stops are not properly ad-
justed.
The paint is damaged.
Electrical defect
Battery is weak or dis-
charged
Activity / Rectification /
Measure
1. Check the maximum net
load of the distributor
module.
2. If necessary, take some
weight off of the module.
1. Contact an authorised
service engineer.
1. Contact an authorised
service engineer.
1. Close valve and deactiv-
ate compressed air sup-
ply system.
2. Contact an authorised
service engineer.
1. Check the maximum net
load of the distributor
module.
2. If necessary, take some
weight off of the module.
1. Contact an authorised
service engineer.
1. Contact an authorised
service engineer.
1. Contact an authorised
service engineer.
1. Replace battery.
Moduevo
IFU 009001001 EN 13