Safety notes
2
Safety notes for the product
20 / 104
D
!
A N GE R
Fire and explosion hazard!
Air, oxygen and oxygen compounds react explosively with oils, greases and
lubricants. The compressed gas is associated with both fire and explosion
hazards.
Always keep the product free of oils, greases and lubricants. Only use sliding
means (lubricants) approved by Getinge for this product.
D
!
A N GE R
Risk of infection!
Splintering paint can fall onto patients and cause infection.
Paint damage must be repaired immediately by Getinge Service or service techni-
cians that have been authorized by Getinge.
W
!
AR N I NG
Risk of injury!
Improper assembly, installation, expansion, adjustment, modification, re-
placement of hard-wired mains connections, maintenance and repair by un-
trained service personnel.
The assembly, installation, extension, correction of settings, change, exchange of
firmly attached wires, maintenance and repair may only be performed by Getinge
service or service technicians that have been authorised by Getinge.
W
!
AR N I NG
Risk of injury as a result of overload!
Changes in product configuration affect the permitted useful load.
The useful load should be recalculated for every change made to the product con-
figuration.
W
!
AR N I NG
Risk of injury!
If the arm or the distributor module are adjusted with force, any unsecured
accessories (e.g. devices on the shelf) may be thrown to the floor.
Do not adjust the distributor module with momentum. If possible, secure accessor-
ies.
W
!
AR N I NG
Risk of collision!
While operating CARGOLIFT or ENERGY with docking unit, if the up & down
buttons are used to adjust the docking system position, there is a risk of col-
lision.
Check that the docking area is totally clear to avoid accidental collisions with other
equipment or human beings.
W
!
AR N I NG
Risk of injury!
While operating CARGOLIFT or ENERGY with docking unit there is a risk of
unexpected accidents due to objects in the docking area, which may lead to
injury and property damage.
Check that there are no objects in the docking area of setting down that might un-
balance the anesthesia machine or Endoscopy trolley.