Getinge MODUEVO Instructions For Use Manual page 17

Medical ceiling supply units

Advertisement

2
Safety notes
2.1
General safety notes
Moduevo
IFU 009001001 EN 13
D
!
A N GE R
Potentially fatal!
Danger resulting from improper handling.
Always observe the instructions for use of the medical ceiling supply units.
D
!
A N GE R
Potentially fatal!
Hazard as a result of unauthorised modifications.
Modifications at the product are not permitted.
The configuration of the product may only be changed by service technicians that
have been authorized by Getinge. In the event of an unauthorized configuration,
the guarantee becomes invalid.
D
!
A N GE R
Potentially fatal!
Electrocution. The ceiling supply units cannot be completely separated from
the power via the On/Off switch on the distributor module.
In order to completely separate the ceiling supply units from the mains voltage (e.g.
in an emergency and for repair and maintenance) use the on-site separation equip-
ment.
D
!
A N GE R
Potentially fatal!
Electrocution. The indicator light might not truly reflect the operational state
of the socket outlet or its supply mains.
Don't determine the status of the supply mains only by the indicator.
D
!
A N GE R
Potentially fatal!
Danger resulting from improper handling.
The integrated or bolt-on rails, which might be located higher than 1000 mm
from finished floor level, can't be used for mounting vacuum collection jars.
D
!
A N GE R
Potentially fatal!
Danger resulting from improper handling.
Always use two hands to disconnect the gas pipe from the gas outlet, one
hand to hold the gas pipe and another hand to press the gas outlet to prevent
the gas pipe from being ejected by the gas or spring, resulting in damage to
the equipment or injuries to personnel.
D
!
A N GE R
Risk of explosion!
The product has no explosion protection and is not suitable for use in AP-M
areas prone to explosion. When using disinfectants that contain alcohol,
cleaning agents or flammable anaesthesia mixtures that are mixed with air,
oxygen or nitrous oxide, explosions may result.
If the product is used in an AP-M area, do not use disinfectants that contain alco-
hol, cleaning agents or flammable anaesthesia mixtures that are mixed with air,
oxygen or nitrous oxide.
Safety notes
General safety notes
17 / 104