Safety notes
2
Safety notes regarding accessories
2.3
Safety notes regarding accessories
22 / 104
D
!
A N GE R
Improper application may result in fatalities!
The description of the operation of components belonging to other manufac-
turers is not part of These instructions for use.
Always observe the instructions for use of the manufacturer!
D
!
A N GE R
Potentially fatal!
Danger resulting from improper handling.
Product information about the Beam Plus Adapter can be found in the instructions
for use for the ceiling supply unit LUCEA 40 and SATELITE XS09. For Information
on operation and installation, please observe these instructions for use.
D
!
A N GE R
Risk of injury!
Devices connected to the interfaces may influence the electrical safety of the
product.
Devices connected to the product's interfaces must comply with the IEC  60601-1
standard and the IEC 60601-1-2 standard.
D
!
A N GE R
Potentially fatal!
In the event of a power failure, devices are only supplied with power if they
are connected to an emergency power supply.
Ensure that all life-saving devices are connected to the emergency power supply.
W
!
AR N I NG
Risk of burns
The high intensity of the light source results in a risk of burns to the eye if
the user looks directly towards the lighthead.
The patient's eyes must be protected during facial surgery. Users must not look dir-
ectly into the light source.
W
!
AR N I NG
Risk of tissue reaction
Light is a form of energy that, on account of certain wavelengths emitted,
may not be suitable for certain pathologies.
The user must be aware of the risks of using the light on subjects who are in-
tolerant to UV and/or infrared light, and on photosensitive subjects.
Before a procedure, please ensure that the light is compatible with this type
of pathology.
W
!
AR N I NG
Risk of drying of tissue or burns
Light is a form of energy that can cause tissue to dry, particularly if light
beams from more than one lighthead are superimposed.
The user must be aware of the risks relating to exposure of open wounds to a
light source of too great an intensity. The user must be vigilant and must ad-
just the level of illumination to suit the patient concerned, particularly during
a lengthy procedure.
Moduevo
IFU 009001001 EN 13