KEYBOARD INTERFACE (NOT AVAILABLE FOR QW2500/
QD2520)

Wedge Quiet Interval

Specifies the amount of time the reader looks for keyboard activity before it breaks the
keyboard connection in order to transmit data to host. The range is from 0 to 990ms in
10ms increments.
1. Determine the desired setting in milliseconds.
2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result
3. Go to
4. Scan the bar code: SELECT WEDGE QUIET INTERVAL SETTING on
5. Scan the appropriate two digits from the keypad in
6. Scan the ENTER/EXIT PROGRAMMING MODE bar code to exit.
This completes the procedure to set the Wedge Quiet Interval. See the table below for
examples of how to set this feature.
Table 13 - Wedge Quiet Interval Setting Examples
STEP
1
Desired Setting
Divide by 10 (and pad with
2
leading zeroes
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT WEDGE QUIET INTERVAL SETTING
Scan two characters from
5
Appendix D
6
Scan ENTER/EXIT PROGRAMMING MODE
NOTE: This feature applies ONLY to the Keyboard Wedge interface.
with leading zeroes to yield two digits. For example: 0 = 00, 5 = 05, 20 = 20, etc.
and scan the ENTER/EXIT PROGRAMMING MODE bar code to enter
page 69
Prog. Mode.
duration which was determined in the steps above. You will hear a two-beep indi-
cation after the last character.
NOTE: If you make a mistake before the last character, scan the CANCEL
bar code to abort and not save the entry string. You can then start again at
the beginning.
ACTION

KEYBOARD INTERFACE (NOT AVAILABLE FOR QW2500/QD2520)

EXAMPLES
10ms
150ms
01
15
'0' and '1'
'1' and '5'
.
page 69
, that represent the
Appendix D
600ms
850ms
60
85
'6' and '0'
'8' and '5'
PRODUCT REFERENCE GUIDE
269