Liquiphant FTL64

Heartbeat Verification

Proof testing for SIL/WHG
1)
devices

Test magnet

Weather protection cover for
dual-compartment housing,
aluminum
1)
Available only for devices with SIL or WHG approval
Endress+Hauser
Heartbeat Monitoring
Continuously provides device and/or process data for an external system. Analysis of this data forms
the basis for process optimization and predictive maintenance.
The "Heartbeat Verification" module includes the Heartbeat Verification Wizard, which performs a
verification of the current device condition and generates the Heartbeat Technology verification
report:
• The wizard can be used via the SmartBlue app.
• The wizard guides the user through the entire process of generating the verification report.
• The operating hours counter and minimum/maximum temperature indicator (peakhold) are
displayed.
• In the event of an increase in the oscillation frequency of the fork, a corrosion warning will appear.
• The order configuration of the oscillation frequency in air is indicated in the verification report. An
increased oscillation frequency indicates the presence of corrosion. A reduced oscillation frequency
indicates that there is buildup present or that the sensor is covered by the medium. Deviations in
the oscillation frequency compared to the oscillation frequency on delivery may occur due to the
process temperature and process pressure.
The "SIL Prooftest", "WHG Prooftest" or "SIL/WHG Prooftest" module includes a wizard for the proof
testing that is required at appropriate intervals for the following applications: SIL (IEC61508/
IEC61511), WHG (German Federal Water Act):
• The wizard can be used via the SmartBlue app.
• The wizard guides the user through the entire process of generating the verification report.
• The verification report can be saved as a PDF file.

Accessories

Order number: 71437508
 51
Test magnet
• Material: stainless steel 316L
• Order number: 71438303
229.4 (9.03)
 52
Weather protection cover for dual-compartment housing, aluminum. Unit of measurement mm (in)
136.7 (5.38)
A0039209
81 (3.19)
103 (4.06)
A0039231