3-3-4 RTC (Real Time Control) Operation
NOTE
1) In the following situations RTC cannot be set.
(1) When TIME (run time) is set to HOLD (continuous operation) in the Run screen
(2) When it is past the start time
(3) When the start time is more than 20 days later than the current date
2) To change the TIME (run time) in the Run screen after setting RTC, cancel RTC and
reset the TIME. To cancel RTC, press the [STOP] button or press the RTC field and
press the [Cancel] button in the RTC setting screen.
3) When logging in as a user (refer to Section 3-2-4 "Logging in as a User"), pressing the
[Cancel] button in the RTC screen will display the PIN entry field shown below.
Enter your login PIN or the administrator PIN in this field.
4) To combine programmed operation (including step-mode operation) and RTC operation,
set RTC after calling the program. The centrifuge totals the run time for all programmed
operation steps to calculate the RTC start time. Programs cannot be called after setting
RTC.
RTC operation is not possible when step-mode operation includes a 0 rpm step.
5) Press the [STOP] or [Cancel] button to stop a run during operation or standby. This
cancels RTC operation and stops the rotor.
Change the run time (centrifugation time) to a numeric value instead of
HOLD.
Set the start time to a time later than the current time.
Set a time to a time that is within 20 days of the current date.