6. Installation
3. Fixing the safety cover
The ultracentrifuge is shipped with the safety cover removed from its rear. When installing the
ultracentrifuge, remove the packing and insert the lower part of the safety cover into the two hooks
protruded from the frame and then secure it with the six M4 binding screws (See Fig.6-2).
M4 binding screw (2 pcs.)
4. Connecting power cord
WARNING : Your ultracentrifuge must be grounded properly.
WARNING : Do not touch the power cord with wet hands to avoid electrical shocks.
All electrical connections should be carried out by a suitable qualified person.
1) When your centrifuge is equipped with a three wire power cord (see below), plug the power cord
coming from the rear of the centrifuge in the terminal of the distribution board in accordance with
ANSI/NFPA 70, NEC, with CSA C22.1, CEC, Part Ⅰ or with both as appropriate. The
green/yellow wire is a grounding wire. Your centrifuge must be grounded properly.
2) When the plug (Part No.84440401) is included in the accessories of your centrifuge and you want
to connect the power cord to the wall receptacle (NEMA 6-30R), connect the plug and the end of
the power cord. For the other connection, follow local electrical codes.
Your centrifuge must be grounded properly.
WARNING : Do not hold the cord when disconnecting the power cord from the receptacle. Hold
the plug when disconnecting it.
Safety cover (shipping status)
Fig.6-2 Securing the safety cover
Black (Neutral)
N
Green/Yellow
(Protective earth)
L
Black (Line)
6-3
Top deck
Hook
M4 binding screws
(2pcs., mounting screws for safety
cover during shipping)
M4 binding screws
(4pcs., accessories)
Safety cover