3-4-6 Admin Function
Step
Touchscreen operation
6
Again enter the PIN entered
in step 5 and press the
[Enter] button.
7
Select an access level.
8
Press the ADMIN tab, the [X]
button, the Run screen tab or
the MENU tab.
NOTE
In the factory defaults, the user name [Admin] is registered as having an access level of [Administrator]
and a PIN of [1111].
Screen displays and notes
An access level setting makes it possible to control access at
three different levels.
The user lockout function has to be enabled to allow user
lockout. For details, refer to Section 3-4-6 (2) "User Lockout."
The following access levels are provided.
Administrator: Allowed to perform any operation.
Supervisor: Allowed to set and change run parameters,
select programs and other operations.
User: Allowed to perform START, STOP and VACUUM
operations.
For details, refer to Fig. 3-4-6 (2) Operations permitted by
each access level.
The added user name and access level are displayed again.
Press the ADMIN tab or the [X] button to return to the ADMIN
screen.
Press the Run screen tab or MENU tab to return to the
respective screen.
3-87
An Access level
dialog box appears.
Access level
dialog box
The User
Management screen
reappears.
Added user