2-2-2 Rotor Chamber

2-2-2 Rotor Chamber
The structure of the rotor chamber (vacuum chamber) is shown in Fig. 2-2-2.
Drive shaft (crown)
Speed sensor
Overspeed detector
(detecting any instance
exceeding the maximum
allowable speed of the rotor)
Protective steeling ring
Note:If sample or water drops to the window of the temperature sensor or the RLM sensor, it may cause an
incorrect detection. Whenever the sensor is wet, wipe it with a clean, dry cloth. Take care not to
scratch the surface of the sensor.
Bowl (Rotor Chamber)
Chamber door seal
Thermoelectric cooling element
(cooling the rotor)
Fig. 2-2-2 Rotor chamber
Handle
Magnetic head for rotor life management (RLM)
sensor (reading and writing information on the life
of the rotor with RLM adapter)
RLM:Rotor Life Managemennt
2-6
Temperature sensor (measuring
the temperature of the rotor)
Chamber door
Door lock system
Vacuum chamber