Step
Touchscreen operation
12
Check the entry and press
the [OK] button.
13
Press the white area in the
ACCEL setting display field
((4)) or the DECEL setting
value display field ((5)) and
set the acceleration and
deceleration mode.
Example)
ACCEL: 5
DECEL: 5
14
To set an RCF value, press
the white area in the
RCFmax setting display field
((6)) or the RCFavg setting
display field ((7)).
NOTE
Operations up to step 16
are only required when an
RCF value is set.
Screen displays and notes
(4)
(5)
Enter the desired code No. and press the [Enter] button.
The [Enter] button is replaced by the [OK] button. Check the
setting and press the [OK] button.
entry/display screen reappears.
For details, see "Setting acceleration and deceleration
mode" in Section 3-2-1 "Setting Run Parameters."
ω
To set an RCF or
respectively.
NOTE
To set an RCF value, the rotor must first be set. Set the flow
as described in step 5. If a rotor has not been set, the
operation will not be accepted.
3-28
3-3-1-1 Program Registration Procedures
The run parameter
entry/display screen
reappears.
The SPEED, TIME,
TEMP display fields and
the on-screen keypad
appear.
Selecting a rotor in step
5 will make it possible to
compute and display
RCFmax and RCFavg
values from the set
SPEED value.
The ACCEL/DECEL
setting screen
appears.
The run parameter
2
value, proceed to step 14 and step 17,
T
The run parameter
entry/display screen
(6)
reappears.
(7)