Press the [START] button.
NOTE
1) Programs cannot be registered, changed or deleted during operation. Perform these
operations when the centrifuge is not running.
2) When the screen is closed by pressing the [X] button in the program display screen
after calling a program, the Function Selection Area appears again but program
operation remains enabled. The [Program] button changes to blue to indicate that the
program is still enabled.
3) If the POWER switch is turned off when the Program display screen is open, the
Program display screen will be closed when the centrifuge is restarted and the
Function Selection Area will appear. However, the program will still remain enabled
and the [Program] button will turn light blue as stated in 2) above.
4) Pressing the [START] button when the [Program] button is lit blue (and a program is
enabled), will start the program and the Function Selection Area is replaced by the
program display screen.
5) To cancel called program operation, call another program or change run parameters.
3-22
3-3-1 Programmed Operation
As the Run screen appears,
the registered run parameters
are set.
program No. is
The
in the [Folder] button in the
Program display screen.
When multiple steps are
registered, the step operation
display indicates the total
number of steps and the
current step.
displayed