1. Specifications
Model
Dimensions
Weight
Power requirement**
Environmental conditions
** The voltage to be used is the one that you specified when purchasing the centrifuge.
The CP100NX/90NX/80NX ultracentrifuges satisfy CE marking requirements. The CE marking is an
international symbol which shows that the product conforms to EC directives.
Standards concerning these directives are as follows:
•
Product safety (EN 61010-1 and EN 61010-2-020)
•
Electromagnetic compatibility (EN 61326-1)
CP100NX
Width: 790 mm; Depth: 690 mm; Height: 880 mm
Depth with safety cover mounted: 890 mm
Height to the handle of the centrifuge door: 925 mm
390 kg
Supply required
Single phase 50/60 Hz
208, 220 Vac+/-10 %, 20 A maximum (normally 8 A)
230, 240 Vac+/-10 %, 16 A maximum (normally 7 A)
Ambient temperature for operation: 2°C to 40°C
•
Ambient Temperature for performance guarantee : 10°C to 30°C
•
Indoor use
•
Altitude up to 2000 m
•
Maximum relative humidity 80 % for temperatures up to 31 °C
•
decreasing linearly to 50 % relative humidity at 40 °C;
Pollution degree: 2
•
Overvoltage category Ⅱ
•
CP90NX
1-2
CP80NX