2-2-1 Touchscreen and External Connection
〔Functions of the Run screen〕
No.
Name and symbol
[VACUUM] button
[START] button
[STOP] button
Function Selection area
[PROGRAM] button
[RCF] button
2
T] button
[RTC] button
The following four stages are displayed, depending on the vacuum of the
rotor chamber.
Atmospheric state.
(1)
The vacuum pump is not active.
Low vacuum. The rotor waits at 4,000 rpm until the
(2)
vacuum reaches medium level.
(3)
Medium vacuum.
(4)
High vacuum.
The orange LED light blinks until the pressure in the rotor chamber equals
the atmosphere pressure.
"Now open the door" is displayed (see below) after the pressure in the
rotor chamber equals the atmosphere pressure.
If the sample is sensitive to temperature increases, press
NOTE
the [START] button when the chamber is at a high vacuum
level.
Press this button to start rotor rotation.
If [VACUUM] is off, pressing this button starts the vacuum pump and
starts temperature control.
Press this button to stop rotor rotation.
This centrifuge incorporates a number of features, such as step-mode
operation and RTC (Real Time Control) operation. Buttons for these features
are displayed and specified in the Function selection area.
Press this button to select the step-mode operation etc (see section 3-3-1).
Press this button to display and set RCF (see section 3-3-2).
Press this button to set ω
Press this button to select the RTC (Real Time Control) operation (see
section 3-3-4).
2-4
Functions and actions
2
T (see section 3-3-3).