Step
Touchscreen operation
Enter the run parameters for
2
step 1 as described in "(1)
Registration Procedures for
Step-Mode Runs."
STEP 1
SPEED: 50,000 rpm
TIME :1 hour
TEMP :15.0 °C
ACCEL: 9
DECEL: 7
Press the [Next Step] button
3
and enter the run
parameters for step 2 from
the run parameter
entry/display screen.
STEP 2
SPEED: 0 rpm
TEMP :15.0 °C
Press the [Next Step] button
4
and enter the run
parameters for step 5 from
step 3.
5
Finally, press the [Register]
button.
Screen displays and notes
Registering a rotor will make it possible to compute and
display RCFmax and RCFavg values from the set SPEED
value.
When a SPEED of 0 rpm is set, TIME is automatically set to
"HOLD."
An arbitrary TIME setting cannot be made.
The entered values are displayed in the STEP 1 to 5 fields.
To correct entry errors, press the STEP field you want to
correct, and press the [Edit] button to make corrections
when the field turns blue.
3-46
3-3-1-3 Step-Mode Operation Procedures
The run parameter
entry/display screen
appears.
[Next Step] button
The run parameter
entry/display screen
in step 2 appears.
STEP No.
[Next Step] button
[Register]
b
The run parameter
setting screen
reappears.
[X] button