Hitachi CP100NX Instruction Manual page 123

Ultracentrifuge
Table of Contents

Advertisement

Step
Touchscreen operation
3
Press the [Edit Runs] button
in the menu box.
From the "Rotor log"
4
check the total number of
runs the rotor was used in the
other centrifuge, enter it and
press the [Enter] button.
If usage data has already
been entered, add the usage
data for the other centrifuge.
A value that is lower than the
currently registered number
of times cannot be entered.
Again select the rotor you
5
want to register, press the
[MENU] button and press the
[Edit Hours] button in the
menu box.
From the "Rotor log"
6
check the total run time the
rotor was used in the other
centrifuge, enter it and press
the [Enter] button.
If usage data has already
been entered, add the usage
data for the other centrifuge.
Use the [ :/ . ] button to enter
decimals.
A value that is lower than the
currently registered time
cannot be entered.
7
Press the Run screen tab to
return to the Run screen.
Screen displays and notes
The display again shows the Run screen.
3-96
3-4-6 Admin Function
The Number of runs
field appears.
The Rotor
management screen
reappears and the
entered number of
runs is displayed in the
total number of runs
indicator in the field of
the selected rotor.
Entered total number
of runs
The Total run time
entry screen
appears.
The Rotor
management screen
reappears and the
entered time is
displayed in the total
run time indicator in
the field of the
selected rotor.
Entered total run time