Hitachi CP100NX Instruction Manual page 69

Ultracentrifuge
Table of Contents

Advertisement

Step
Touchscreen operation
1
As described in steps 1 to 8
in "(1) Procedures for
registration of run
parameters" in Section
3-3-1-1 "Program
registration procedures,"
enter the Program No.,
registered rotor, user and
remarks, select a folder color
and press the [Next] button.
In this example, the steps are
stored in program No. 10.
2
Press a field in Step 1, check
that the field turns blue and
press the [Edit] button.
Enter the run parameters for
3
step 1 from the run
parameter entry/display
screen.
STEP 1
SPEED: 100,000 rpm
TIME :1 h 30 min
TEMP :4.0 °C
ACCEL: 9
DECEL: 9
Press the [Next Step] button
4
and enter the run parameters
for step 2 from the run
parameter entry/display
screen for step 2.
STEP 2
SPEED: 80,000 rpm
TIME : 1 h
TEMP :4.0 °C
ACCEL: 9
DECEL: 9
Screen displays and notes
 Program registration screen
[Next] button
Registering a rotor will make it possible to compute and
display RCFmax and RCFavg values from the set SPEED
value.
The value entered in STEP 1 is copied to the run parameter
entry/display screen.
3-42
3-3-1-3 Step-Mode Operation Procedures
 Run parameter setting/display
[Edit] button
The run parameter
entry/display screen
appears.
[Next Step] button
The run parameter
entry/display screen
appears.
[Next Step] button
The run parameter
entry/display screen
in STEP 2 appears.
Entry screen STEP No.

Advertisement