Hitachi CP100NX Instruction Manual page 68

Ultracentrifuge
Table of Contents

Advertisement

3-3-1-3 Step-Mode Operation Procedures
3-3-1-3 Step-Mode Operation Procedures
This centrifuge is provided with a "step-mode operation" function that can store two or more run
parameters in one program memory area and a total of 30 steps. This allows it to change speed, run
time, temperature and other conditions during operation.
For example, you could set the speed of a step to 0 rpm to stop rotation to enable opening and closing
vacuum leaks and doors. This makes it possible to register in advance the repeat count of samples
that are removed and processed each run under the same separation conditions, which prevents
processing samples too many or too few times.
Operation procedures are described using examples below.
(1) Registration Procedures for Step-Mode Operation
[Setting example]
The following describes the setup procedures for continuous operation under the three-step run
parameters shown in the figure below.
In this example, the operations described in Fig. 3-3-1 (1) will be performed automatically.
Speed
Run time
Temperature
Acceleration
mode
Deceleration
mode
100,000
Speed
80,000
(rpm)
60,000
Step 1
100,000 rpm
1h30min
°C
4.0
9
9
Step 1
1h 30min
Fig. 3-3-1 (1) Example of Step-Mode Operations
3-41
Step 2
80,000 rpm
1h
°C
4.0
9
9
Step 2
1h
30 min
Step 3
60,000 rpm
30min
°C
4.0
9
7
Step 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cp90nxCp80nx

Table of Contents