 Setting speed, run time and temperature
Operation procedures will be described using examples.
Setting Item
Setting example
1
Press the SPEED,
TIME or TEMP field
to display the
on-screen keypad.
2
Each field displays
the value set for the
previous run. If the
color of the initial
digits of the setting
are white, again
press the field of the
item you want to set.
If the color of the
initial digits are blue,
go to step 3.
3
Press the on-screen
keypad to enter a
numeric value.
7
8
4
5
1
2
0
Each new digit that
is entered is scrolled
to the left.
4
Check the entered
value. To continue
making other settings,
press the field of the
desired setting. When
complete, press the
[Enter] button.
Use the [CE] button
to cancel incorrect
entries.
Setting range and units
Speed
(SPEED)
100,000 rpm
Color: Blue
1
0
The last two digits
are fixed.
×
9
CE
6
: / .
3
Hold
Enter
Set it to 100,000
rpm.
Set a value between
1,000 rpm to
maximum speed in
100 rpm increments.
run time
(TIME)
2 hours 30 minutes
Color: Blue
:
2
: /.
0
0
Press the [ : / . ]
button to move the
cursor to the
"minutes" position.
To start a
continuous run,
press the [Hold]
button.
Set it to 2:30.
(2 hours 30 minutes)
Set a value up to 999
hours and 59
minutes in 1 minute
increments.
3-8
3-2-1 Setting Run Parameters
Temperature
(TEMP)
°C
4
Color: Blue
3
0
4
Set it to 4
Temperatures in the
°C
0 to 40
range can
be set in 0.1℃
increments.
°C
.