3-3-1-1 Program Registration Procedures
Step
Touchscreen operation
15
Press the display area of the
item you want to set and
enter the desired value.
Example)
RCFmax:513,700×g
16
Check the set value and
value computed from the set
value and press the [OK]
button.
To set a ω
2
T value, press the
17
white area in the ω
value display field. ((8)).
NOTE
Operations up to step 18
are only required when a
ω
2
T value is set.
When an RCFmax or RCFavg value is entered, the other
value is automatically computed and displayed.
Enter the desired value and press the [Enter] button.
The SPEED value computed from the set RCF value and
rotor information is displayed in the SPEED display.
NOTE
If the SPEED value entered in step 11 differs from the
SPEED value computed from the RCF value set here, the
value entered in step 11 is replaced with the computed value.
2
T setting
NOTE
To set a
is not accepted if SPEED is not set.
Screen displays and notes
(8)
ω
2
value, SPEED must first be set.
T
3-29
The RCF setting screen
appears.
Set rotor model
SPEED display
RCFavg setting value
display field
RCFmax setting value
display field
The RCFavg value is
computed and displayed
from the set RCFmax
value.
SPEED display
[OK] Button
The run parameter
entry/display screen
reappears.
This operation