Hitachi CP100NX Instruction Manual page 110

Ultracentrifuge
Table of Contents

Advertisement

3-4-5 Customizing
Step
Touchscreen operation
5
Press the "CUSTOM" tab, the
[X] button, the Run screen
tab or the MENU tab.
NOTE
1) Economy mode is enabled only when the rotor is at standstill and the vacuum pump is not
operating.
2) During economy mode, the backlight on the touchscreen is dimmed, the cooling system is
stopped and no electricity is supplied to release the door lock. If the door is locked when the
centrifuge enters the economy mode, the door is locked and cannot be opened.
3) To temporarily cancel economy mode, touch the touchscreen. This will enable all operations and
the door can be opened.
4) Even if economy mode is temporarily canceled, the centrifuge will reengage economy mode
when the set wait time elapses as long as the rotor is at standstill and the vacuum pump is not
operating.
5) When room temperature is high and the instrument needs to be cooled, the cooling fan may
continue operating even when economy mode is engaged.
6) The waiting time until economy mode is activated can be set to a time between 1 and 180
minutes.
7) Since the touchscreen backlight is dimmed in economy mode, it may be difficult to realize that
the instrument is on. So be sure to turn it off when you have completed a run.
8) The status display lamp (LED indicator) can be set to continue operation in economy mode to
make it easier to see that the instrument is powered. For details, refer to (8) "Status display lamp
(LED indicator) setting."
Instrument operations and notes
Press the CUSTOM tab or the [X] button to return to the
CUSTOM screen.
Press the Run screen tab or MENU tab to return to the
respective screen.
To disable economy mode, press the [Auto-standby] button
and the [Disable] button.
3-83

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cp90nxCp80nx