Basic Operation; Setting Run Parameters - Hitachi CP100NX Instruction Manual

Ultracentrifuge
Table of Contents

Advertisement

3-2 Basic Operation

3-2-1 Setting Run Parameters

WARNING: (1) Do not incline or move the instrument while the rotor is spinning. Do not place any
(2) The centrifuge itself may move if the rotor fails during high-speed rotation. Make
CAUTION: (1) Do not press the touchscreen with a sharp-pointed object such as a ballpoint pen.
(2) If abnormal sound or vibration occurs, stop the operation immediately and contact
3-2-1 Setting Run Parameters
This section will describe the Run screen, the starting point of basic operations. For information on
screens displayed during normal operation and entry of run parameters, refer to Section 2-2-1
"Touchscreen and External Connections."
[Normal operation display]
The screen for displaying run parameters and operating conditions is called the run screen.
SPEED, TIME and TEMP are displayed in two rows: the top row displays the current operating
condition, while the bottom row displays the setting value.
The acceleration (ACCEL) and deceleration (DECEL) fields display setting values.
Acceleration/
deceleration
display fields
3-2 Basic Operation
object on the instrument or lean on the instrument.
sure there is a 30cm area around the centrifuge that will allow for such movement
and do not allow individuals to enter into that area during operation. Also do not
place dangerous objects such as flammable or explosive materials on top of the
centrifuge or in the surrounding area.
a Hitachi Koki authorized sales/service representative.
Fig. 3-2-1 Run screen
3-5
Operating condition
display field
Setting value
display field

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cp90nxCp80nx

Table of Contents