Hitachi CP100NX Instruction Manual page 70

Ultracentrifuge
Table of Contents

Advertisement

3-3-1-3 Step-Mode Operation Procedures
Step
Touchscreen operation
5
Press the [Next Step] button
and enter the run
parameters for step 3 from
the run parameter
entry/display screen for step
3.
STEP 3
SPEED: 60,000 rpm
TIME :30 min
TEMP :4.0 °C
ACCEL: 9
DECEL: 7
6
Press the [Register] button.
7
Check the entries displayed
for each step in the run
parameter setting/display
screen. If they are correct,
press the [X] button and
press the [X] button in the
Program registration screen.
8
Check that the folder you
just registered has been
added to and is displayed in
the Program list screen
before pressing the [X]
button.
Screen displays and notes
Like in STEP 1, the value entered in STEP 2 is copied to the
run parameter entry/display screen.
The entered values are displayed in the STEP 1, 2 and 3
fields.
To correct entry errors, press the STEP field you want to
correct, and press the [Edit] button when the field turns blue.
The Run screen reappears.
3-43
The run parameter
entry/display screen
in STEP 3 appears.
[Register] button
The run parameter
setting/display screen
reappears.
[X] button
The Program list
screen reappears.
[X] button
The program
registered in this
example

Advertisement