Breathing System Specifications - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Carestation
620/650/650c (A1)

Breathing system specifications

11-12
Compressible volume Ventilator side
Bag side
Reusable canister
Disposable canister
Absorbent
Reusable canister
Disposable canister
Connections
Auxiliary Common Gas Outlet: ISO 5356 type
connector on the front of the system (standard 22
mm OD or 15 mm ID conical friction fit connectors).
Breathing system
Less than or equal to 150 ml/min total at 3 kPa (30
leakage
cmH2O) (both in absorber mode and with canister
removed).
Breathing system
Volume of gas lost due to internal compliance of
compliance
filled reusable absorber canister (bag mode only):
1.74 ml/0.098 kPa (1 cmH2O)
52.4 ml/3 kPa (30 cmH2O)
Volume of gas lost due to internal compliance of
filled disposable absorber canister (bag mode only):
1.81 ml/0.098 kPa (1 cmH2O)
54.56 ml/3 kPa (30 cmH2O)
APL valve
Approximately 0 to 70 cmH2O
Negative pressure
Minimum 100 ml/min at -14 cmH2O
relief
Mechanical
115 +/- 10 cmH2O
overpressure valve
Inspiratory breathing resistance in bag mode*
l/min
kPa
Absorber canister installed
5
0.030
30
0.172
60
0.456
Absorber canister removed
5
0.034
30
0.159
60
0.401
*Values include patient circuit tubing and Y-piece 0.065 kPa (0.65 cmH2O)
at 60 l/min. Patient circuit tubing and breathing system configurations affect
resistance.
2006 ml
500 ml
1000 ml
1000 ml
1370 ml
1400 ml
cmH2O
0.30
1.72
4.56
0.34
1.59
4.01
2076152-001