Carestation
620/650/650c (A1)
Turning on the system
CAUTION
3-4
1.
Plug the power cord into an electrical outlet. Make sure the
mains indicator light is on.
•
The mains indicator is lit when AC power is connected.
•
Battery is charging if it is not already fully charged.
1.
Mains indicator
2.
On/Standby switch
Figure 3-1 • Mains indicator and On/Standby switch
2.
Check that the breathing system is properly connected.
Do not turn on the system with the right-hand
(inspiratory) port plugged.
3.
Push the On/Standby switch for 1 second to turn on the system.
The Total flow indicator shows total fresh gas flow and the
display shows the power-up screen.
The system does a series of automated self tests.
In case of emergency, during power-up, O2 and agent delivery is
possible in bag mode.
4.
Perform a Full Test before the first case of the day.
5.
Perform a preoperative checkout before each case. See the
"Preoperative checkout" section.
Note
The system must perform a power-up self test after 12 hours of
remaining on. If the system has been on longer than 12 hours
without a power-up self test, the 'Turn power Off and On for self
tests' alarm occurs. Turn the power off and then back on between
cases to resolve the alarm.
1
2
2076152-001