Setting MV/TV alarms

Setting alarm limits

WARNING

Viewing alarm history

3-22
Use the MV/TV Alarms setting to turn off the MV and TV alarms.
Settings made during manual ventilation are not retained when
mechanical ventilation starts. Settings made during mechanical
ventilation are retained when manual ventilation starts.
For example, if MV/TV Alarms is set to Off during manual
ventilation, the alarms remain off until the Bag/Vent switch is set to
Vent or the MV/TV Alarms is set to On.
For example, if the MV/TV Alarms is set to Off during mechanical
ventilation, the alarms remain off when manual ventilation starts.
1.
Select Alarm Setup.
2.
To turn the volume alarms off, set MV/TV Alarms to Off.
'MV/TV Alarms Off' appears in the general message field.
The volume alarm limits waveform numerics shows alarm off
symbol during a case.
3.
To turn the volume alarms on, set MV/TV Alarms to On.
4.
Select Close.
Do not set alarm limits to extreme values. Setting limits to
extreme values can render the alarm useless.
1.
Select Alarm Setup.
2.
From the Primary Limits and More Limits tabs, select the
alarm limit and make the change.
3.
Push the Home key, touch the waveform area of the display, or
select Close to close the menu.
Use the Alarm History tab to view the list of the 12 most recent high
and medium priority alarms that occurred since the start of the case.
The alarm history clears at the start of a new case.
1.
Select Alarm Setup.
2.
Select the Alarm History tab.
The list of alarms shows in the window.