Alarm ranges

2076152-001
The alarm names are listed in the Primary Limits and More Limits
tabs on the Alarm Setup menu. See the "Super user mode" section
for more information on the alarm default settings.
Alarm
Range
Pmax (only high)
12-100 cmH2O
1.2 - 9.8 kPa
12 - 98 mbar, hPa
9-73 mmHg
MV High
0.5 - 30.0, Off l/min
MV Low
Off, 0.1 - 10.0 l/min
TV High
20 -1600, Off ml
TV Low
Off, 1 - 5 ml
5 - 20 ml
20 - 1500 ml
RR High
2 - 100, Off /min
RR Low
Off, 1 -99 /min
EtCO2 High
0.1 - 15%, Off
0.1 - 15, Off kPa
1 - 115, Off mmHg
EtCO2 Low
Off, 0.1 - 14.9%
Off, 0.1 - 14.9 kPa
Off, 1 - 114 mmHg
FiCO2 High
0.1 - 15%, Off
0.1 - 15, Off kPa
1 - 115, Off mmHg
FiO2 High
19 - 100%, Off
FiO2 Low
18 - 99%
EtO2 High
19 - 100%, Off
EtO2 Low
Off, 1 - 99%
FiIso High
0.1 - 7.0%
FiIso Low
Off, 0.1 - 6.9%
FiSev High
0.1 - 10.0%
FiSev Low
Off, 0.1 - 9.9%
FiDes High
0.1 - 20.0%
FiDes Low
Off, 0.1 - 19.9%
FiEnf High
0.1 - 7.0%
FiEnf Low
Off, 0.1 - 6.9%
FiHal High
0.1 - 7.0%
FiHal Low
Off, 0.1 - 6.9%
EtIso High
0.1 - 7.0%, Off
7 Alarms and troubleshooting
Increment
1 cmH2O
0.1 kPa
1 mbar, 1 hPa
1 mmHg
0.5 l/min
0.1 l/min
20 ml
1 ml
5 ml
20 ml
1 /min
1 /min
0.1%
0.1 kPa
1 mmHg
0.1%
0.1 kPa
1 mmHg
0.1%
0.1 kPa
1 mmHg
1%
1%
1%
1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
0.1%
7-13