Connecting the Pedi-lite and D-lite sensors
WARNING
2076152-001
Use the Pedi-lite sensor for patients with tidal volumes up
to, and including 200 ml.
™
•
Use the D-lite
greater than 200 ml.
•
To ensure patient safety, make sure that the gas
sampling connectors are connected as described in
these instructions and are not interchanged with
connectors on other equipment.
The D-lite and Pedi-lite sensors have a port to provide gas samples
to the airway module. Both sensors are available as reusable
(yellow) or disposable (clear).
Place all D-lite ports upwards with a 25 to 35° tilt to prevent
condensed water from entering the sensor interior and the tubings.
Use manufacturer approved gas sampling lines (PVC/ PE) when
anesthetic agents are used.
1.
Connect the spirometry tubes from the Pedi-lite or D-lite sensor
to the airway module.
2.
Connect the gas sampling line from the Pedi-lite or D-lite sensor
to the D-fend Pro water trap on the airway module.
1
2
1.
Airway module
2.
Gas sampling line connector on the water trap
3.
Gas sampling and spirometry tubes
4.
Gas sampling line
5.
Pedi-lite or D-lite sensor
Figure 6-4 • Airway gas connection to airway module with spirometry
sensor for patients with tidal volumes
3
6 Airway modules
4 5
6-7