ecoFLOW

1.
Agent
2.
Agent cost
3.
Agent flow
4.
FiO2 flow marker
5.
FiO2 flow
6.
O2 total
7.
N2O total
8.
N2 total
9.
Flow bobbin
10. Total flow
Figure 3-3 • ecoFLOW feature
2076152-001
This feature provides a split screen view that shows the approximate
minimum O2 flow to maintain a preset inspired O2 concentration.
Also shown is the approximate agent used per hour and the cost.
The ecoFLOW functionality is only available when an airway module
is installed and the system is delivering mechanical ventilation.
20
0
-20
10
Total Flow
l/min
15
9
9
3
8
2
7
1
6
O2 Total
0.99
Fi25 O2 flow
5
Shows Agent cost and flow information.
The cost of the current agent flow. This value is determined by the agent flow multiplied
by the agent cost set in Super user mode.
The measured value of the liquid agent flow from the vaporizer. The agent flow may have
a delayed response. For example: Iso.
The graphical representation on the flow tube of the FiO2 flow value. This marker can be
removed by disabling it in Super user mode.
The minimum O2 flow needed to maintain the set inspired O2 flow. This item can be
disabled in Super user mode. For example: Fi25 O2 flow.
O2 total equals the set of O2 flow. If N2O is the balance gas, this equals the set O2 flow.
If Air is the balance gas, this is the set of O2 flow plus 21% of the Air flow.
N2O total equals the set N2O flow.
N2 total equals the N2 in the Air flow.
The height of this represents the total fresh gas flow delivered to the breathing system.
Shows Total Flow information.
40
60
80
100
Agent
1
21.91
2
ml/h
500
50
Iso
3
4
0.60
4
3 Operation
3-35