Setting spirometry volume type

Saving, viewing, and deleting spirometry loops

2076152-001
1.
Select System Setup - Screen Setup - More Settings or
Spirometry - Setup Loops.
2.
Select Data Source.
3.
Select Patient to have spirometry data sourced from the airway
module or Vent to have spirometry data sourced from the
ventilator.
Note
When using the Airway Gas Option, only Vent is available for the
Data Source.
Note
If no airway module is installed, all settings will default back to Vent.
4.
Select Back to view changes made and access other functions
of the Spirometry menu.
The volume shown on the spirometry split screen can be set to
minute volume or tidal volume.
1.
Select Spirometry - Setup Loops.
2.
Select Show MV or TV and select MV or TV from the drop-
down list.
Set to TV to show TVinsp and TVexp on the spirometry split
screen.
Set to MV to show MVexp and TVexp on the spirometry split
screen.
3.
Select Back to view changes made and access other functions
of the Spirometry menu.
Spirometry loops can be saved, viewed, and deleted through the
Spirometry menu.
1.
Select Spirometry.
2.
To store a loop to memory, select Save Loop. Up to six loops
can be saved.
3.
To view a saved loop, set Show Ref. Loop to the time at which
it was saved.
4.
To delete a saved loop, set Delete Ref. Loop to the time at
which it was saved.
5.
Select Close.
3 Operation
3-29