Carestation
620/650/650c (A1)
Connecting the airway module
WARNING
6-6
1.
Check that the airway module is installed.
2.
Check that the airway adapter connections are tight and that the
adapter is correctly installed.
3.
Check that the water trap container is empty and properly
attached.
Empty the water trap container whenever the container is more
than half full. Under normal conditions, the D-fend Pro container
fills in 24 hours.
Before connecting the exhaust line to the sample gas
outlet on the airway module, ensure the other end is
connected to the sample gas return port on the
anesthesia machine. Incorrect connections may cause
patient injury.
4.
Connect a line from the airway module sample gas outlet to the
sample gas return port on the anesthesia machine.
5.
Attach the gas sampling line to the sampling line connector on
the water trap.
6.
Turn the system on. The system does a series of automated self
tests. The automatic agent identification is activated.
7.
Connect the sampling line to the airway adapter. Take the gas
sample as close to the patient's airway as possible. Position the
adapter's sampling port upwards to prevent condensed water
from entering the sampling line.
1
1.
Airway module
2.
Gas sampling line
3.
Airway adapter with sampling line connector
Figure 6-3 • Airway gases setup with airway module
2
3
2076152-001