GE Carestation 620 User's Reference Manual page 244

Table of Contents

Advertisement

Carestation
620/650/650c (A1)
Guidance and manufacturer's declaration - electromagnetic
immunity
Phenomenon
Electrostatic discharge
Radiated RF EM fields
Proximity fields from RF
wireless communications
equipment
Rated power frequency
magnetic fields
Phenomenon
Electrostatic fast transients /
bursts
Surges
Line-to-line
Surges
Line-to-ground
Conducted disturbances
induced by RF fields
Voltage dips
Voltage interruptions
11-42
The system is suitable for use in the specified electromagnetic
environment. The customer and/or user of the system should assure
that it is used in an electromagnetic environment as described below.
.
Basic EMC
Professional healthcare facility environment
standard or test
method
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4 kV, +/- 8 kV, +/- 15 kV air
IEC 61000-4-3
3 V/m
80 MHz - 2.7 GHz
80 % AM at 2 Hz (test performed at 2 Hz is worst case as
identified for risk management)
(system compliant to 10 V/m immunity test level)
IEC 61000-4-3
See the
RF wireless communication equipment
table in "Recommended separation distances".
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
.
Basic EMC
Professional healthcare facility environment
standard or test
method
IEC 61000-4-4
+/- 2 kV
100 KHz repetition frequency
IEC 61000-4-5
+/- 0.5 kV, +/- 1 kV
IEC 61000-4-5
+/- 0.5 kV, +/- 1 kV, +/- 2 kV
IEC 61000-4-6
3 V
0.15 MHz - 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
80 % AM at 2 Hz (test performed at 2 Hz is worst case as
identified for risk management)
(system compliant to 10 V immunity test level)
IEC 61000-4-11
0 % UT: 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and
315°
0 % UT: 1 cycle and 70 % Ut: 25/30 cycles
sine phase at 0°
IEC 61000-4-11
0 % UT: 250/300 cycle
2076152-001