Carestation
620/650/650c (A1)
7-16
9.
Test the Ppeak low. Leak? alarm:
•
Unblock the patient connection.
•
Set the Bag/Vent switch to Vent.
•
Set the tidal volume and total flow to minimum.
•
Other alarms such as MVexp low can occur.
•
Make sure that the Ppeak low. Leak? alarm occurs.
10. Set all alarm limits to approved clinical values.
2076152-001