System Utility Function Setting
DHCP
If "DHCP" is selected, the DHCP server in the current network will assign the
network parameters (e.g. IP address, Subnet, Gateway, and DNS) for the
instrument.
Auto IP
When "Auto IP" is selected, the instrument will acquire the IP address ranging
from "169.254.0.1" to "169.254.255.254" and the subnet mask (255.255.0.0)
automatically based on the current network configuration. The "Auto IP" works
only when "DHCP" is not selected or connection is failed.
Static IP
If "Static IP" is selected, the instrument is configured with static IP. In this case,
you need to disable DHCP and Auto IP manually. Then you need to configure the
parameters such as "IP address", "Subnet", "Gateway", and "DNS" manually. At
this time, you can self-define the network parameters (e.g. IP address) of the
instrument.
-
-
-
-
DG900 Pro User Guide
98
Set the IP address
The format of the IP address is nnn.nnn.nnn.nnn. The range of the first
segment (nnn) of the address is from 0 to 255 (except 127); wherein, the
valid range is from 0 to 223. The range for the other three segments is from
0 to 255. You are recommended to ask your network administrator for an IP
address available.
Set the subnet mask
The format of the subnet mask is nnn.nnn.nnn.nnn. Wherein, the range of
"nnn" is from 0 to 255. You are recommended to ask your network
administrator for a subnet mask available.
Set the default gateway
You can set this parameter in Static IP mode. The format of the gateway is
nnn.nnn.nnn.nnn. The range of the first segment (nnn) is from 0 to 223
(except 127), and the range for the other three segments is from 0 to 255.
You are recommended to ask your network administrator for a gate address
available.
Set the DNS address
You can set this parameter in Static IP mode. The format of the DNS address
is "nnn.nnn.nnn.nnn". The range for the first segment (nnn) of the address is
from 0 to 223 (except 127); and the range for the other three segments is
from 0 to 255. You are recommended to ask your network administrator for
an address available.
Generally, you do not need to set the DNS, therefore this parameter setting
can be ignored.