ABB ProKiln GAC400 Operating Instruction page 61

Probe gas sampling system
Table of Contents

Advertisement

ProKiln GAC400 PROBE GAS SAMPLING SYSTEM | OI/GAC400-EN REV. A
7. Reset all alarms and maintenance requests
Home page ►Alarm Reset
Push the Emergency Stop Reset
control cabinet
Turn the Operation/Service switch
5 Emergency Stop Reset
Figure 50: Control cabinet
8. Perform the following checks to ensure smooth operation of
the probe gas sampling system before inserting the probe
for the first time:
After switching on: Working on the cooling module on
page 62
Checking the pneumatic system on page 62
Starting frequency converter and adapting control
parameters on page 63
Checking/correcting operating parameters on page 64
5
on the front of the
6
to Operation Mode
5
6
6 Operation/Service switch
Power breakers in the control cabinet
The illustration below shows the power breakers in the control
cabinet.
1
2
Pos. Component
1
M1 Retractor drive motor
2
M1 Retractor position brake
3
Water system power supply
4
H1 Probe sample tube heater
5
H2 Sample hose heater
6
H3 Heated sample shut-off valve (optional)
7
Control cabinet internal power outlet
8
Control power supply 24 V DC
*
Listed rating is included as an example as the as build EL-
DOCUMENTATION FOR PROBE SYSTEM enclosed with the delivery of the
system will have the supplied rating for a specific customer.
8
7
4
3
5
6
Rating*
16 A
2 A
10 A
10 A
6 A
6 A
10 A
10 A
61