ProKiln GAC400 PROBE GAS SAMPLING SYSTEM | OI/GAC400-EN REV. A
Parameter input Alarm setpoint menu
Figure 29: Alarm setpoint menu
Parameter input
Function
T01 TAH
Heated probe temperature high
maintenance request
T01 TAL
Heated probe temperature low
maintenance request
T02 TAH
Heated hose temperature high
maintenance request
T02 TAL
Heated hose temperature low
maintenance request
T03 TAH
Heated shut-off valve temperature
high maintenance request
T03 TAL
Heated shut-off valve temperature low
maintenance request
T10 TAHH
Water tank temperature high-high
alarm
T10 TAL
Water tank temperature low
maintenance request
T11 TAHH
Water temperature high-high forward
cooling water flow alarm
T12 TAHH
Water temperature high-high return
cooling water flow alarm
F12 FAL
Water flow return low maintenance
request
F12 FALL
Water flow return low-low alarm
T12-T11 TAHH
Temperature difference (forward-
return) high-high alarm
Default value
200 °C
170 °C
190 °C
150 °C
190 °C
150 °C
80 °C
5 °C
80 °C
90 °C
80 l/m
60 l/m
25 °C
33