ProKiln GAC400 PROBE GAS SAMPLING SYSTEM | OI/GAC400-EN REV. A
... 11 Maintenance
... Maintenance work
Instructions: checking probe vertical and horizontal position
How to check probe position:
1. Turn the Safety switch -101S2 at the probe rear cabinet to
OFF position, to prevent any movement of the probe. This
switch can also be secured with a padlock.
2. With the probe inserted, check the distance between the
flange and the tube sealing device. Target requirement: The
flange must be touching
3. If this target requirement is not met, adjust the probe by
means of the following actions:
Adjust the probe flange (see step 4)
Adjust vertical position (see step 5)
Adjust horizontal position (see step 5)
4. Adjusting the probe flange:
Release bolt
to enable the probe flange (Green) to
1
move forth or back as required.
When the probe is in sample position S3 of the retractor,
adjust the probe flange to go firmly against the counter
flange
2
on the flange shutter.
2
1 Bolt
Figure 56: Adjusting the probe flange
2 Counter flange
5. Adjusting the vertical position:
Release the support bolt
allow for adjustments
Release the 3 adjustment bolts on each side of the
retractor buggy base
adjustments.
Adjust vertical position so that sealing flange is aligned
with counter flange.
When done tighten the bolts min 90 Nm.
Adjust support bolt
the probe.
Adjusting the horizontal position:
Release the support bolt
allow for adjustments
Release the 2 adjustment bolts on each side of the Probe
Mounting plate
Adjust horizontal position so that sealing flange is
aligned with counter flange
When done tighten the bolts min 90 Nm.
Adjust support bolt
the probe.
1
1
2
3
1 Support bolt
2 Adjustment bolts
Figure 57: Adjusting the vertical and horizontal position of the probe
6. Turn the Safety switch -101S2 to ON, to allow normal
operation of the probe.
1
(red) supporting the probe to
2
(blue) for height and angle
enough so it provides support for
1
1
(red) supporting the probe to
3
(green) for angle adjustments.
1
enough so it provides support for
3 Mounting plate

Advertisement

Table of Contents
loading

Table of Contents