ProKiln GAC400 PROBE GAS SAMPLING SYSTEM | OI/GAC400-EN REV. A
5. Perform maintenance tasks as required, always observing
the operation manual.
6. Leave danger area around the retractor and close the
protective installation. Check surroundings and make sure
that no-one is standing in the area of probe travel before
proceeding with next step.
7. De-actuate the EMERGENCY-STOP switch and turn the Safety
switch -101S2 at the probe rear cabinet to OFF
8. Push Emergency-Reset button -15SB/-15P12 at the front of
the control cabinet to allow movement of the probe.
9. Turn Service/Operation mode switch -S50S15 clockwise to
the Operation Mode position.
10. Re-insert the probe using the HMI at the control cabinet.
Path: Home ► Manual ► Start to S3
WARNING! The probe moves into the kiln. Do not stand in the
area of travel of the probe or probe buggy.
11. When probe has reached sampling position S3 enter Auto
Mode operation
Path: Home ► Auto
101