Making Relative Measurements; Procedure - Agilent Technologies E4419B User Manual

Dual channel power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Making Relative Measurements

Relative mode allows you to compare any measurement result to a
reference value. The relative reading can be displayed in either dB or % by
dBm/W
pressing
and selecting either
result is displayed in % a prefix multiplier may be shown.
Relative mode is window dependent and when enabled displays "Rel" in
the window to which it's applied.

Procedure

To set a reference value on the currently selected window:
Rel
1. Press
.
Offset
2. Press
to use the current reading as the reference value. This
Rel
allows you to compare any measurement result in dB or
percentage (%).
3.
Rel Off On
Successive measurements are now displayed relative to the reference
value. The relative mode can be disabled and re-enabled simply by
pressing
Rel Off On
Agilent E4419B User's Guide
or
dB
is automatically set to "On" when
.
Power Meter Operation
Making Relative Measurements
. When the measurement
%
is pressed.
Rel
2-37

Hide quick links:

Advertisement

Table of Contents
loading