Replacing the backup battery

The backup battery can be replaced without turning off the unit. Make sure that the board
does not short-circuit and that the soldering iron is not connected to main power while sol-
dering.
Read out the exposure counter, refer to the section "P16 - Show the exposure counter".
Read out the mAs limit, refer to the section "P15 - upper mAs limit".
Read out the kV step and limit, refer to the section "P14 - Number of kV steps and upper
kV limit".
Loosen the screws to the system upper cover and remove it.
Loosen the screws and remove the protection plate.
Locate the battery, refer to the wiring diagram, page 110.
To avoid damage to components or assemblies, ensure compliance with ESD
guidelines.
Remove the battery jumper.
Desolder the battery and connect the new battery with the soldering iron.
Reattach the battery jumper.
Reattach the protection plate.
Switch off the unit for 20 s.