Follow these steps
Charged Aerosol Detectors (VH-D20, VF-D20)
Operating Manual
No.
Description
5
USB hub ("A" type connector)
Allows connection to other modules in the Vanquish system
6
USB (Universal Serial Bus) port ("B" type connector)
Allows connection to other modules in the Vanquish system or the
computer on which the data management system is installed, such as the
Chromeleon software
7
Rating plate, indicating the serial number, part number, module name,
revision number (if any), and the line and fuse rating
8
Gas inlet
Allows connection to a gas supply
9
Gas exhaust
Allows connection to ventilation
TIP Thermo Fisher Scientific recommends using the USB ports only as
described above. If the USB ports are used for any other purpose,
Thermo Fisher Scientific cannot ensure proper functionality.
NOTICE
Never use defective communication cables. If you suspect that a
cable is defective, replace the cable.
To ensure trouble-free operation, use only the cables provided by
Thermo Fisher Scientific for connecting the device.
1. Place the detector in the system as required by the system
configuration. See
2. Connect the required interface cables to the detector. See
Connecting the Interface Cables
3. Connect the power cord (see
(} page 62)).
System Arrangement
(} page 56).
(} page 60).
Connecting the Power Cord
5
Installation
Page 59