Thermo Scientific VH-D20 Operating Manual page 57

Charged aerosol detectors
Hide thumbs Also See for VH-D20:
Table of Contents

Advertisement

Follow these steps
Charged Aerosol Detectors (VH-D20, VF-D20)
Operating Manual
No.
Description
4
Autosampler
5
Pump
6
System Base
7
Column Compartment
Depending on the system configuration, follow the steps as required:
System configuration
If the charged aerosol detector
is the only detector in the
Vanquish system
If the charged aerosol detector
is the second detector in the
Vanquish system, for example
after a diode array detector
Steps
Follow the instructions in the Vanquish System
Operating Manual to set up the system stack.
1. Set up the system stack with the first detector in
the system, such as the diode array detector. For
instructions, refer to the Vanquish System
Operating Manual. Do not install the solvent rack
and the column compartment yet.
2. To position the charged aerosol detector, lift it
by its carrying handles. Place the detector on top
of the stack approximately 5 cm before the end
of the rails. Lift the front part of the detector and
push the detector towards the rear until the
detector clicks into place.
3. To remove the carrying handles, follow the
corresponding steps in
Unpacking
4. Place the solvent rack on the detector
approximately 5 cm before the end of the rails.
Lift the front part of the solvent rack and push
the solvent rack towards the rear until it clicks
into place.
Note: The charged aerosol detector needs to be
at the end of the flow path because it is a
destructive detector.
5. Install the column compartment. Refer to the
Operating Manual for the column compartment.
5
Installation
(} page 41).